How to Calibrate a Hanna Salinity Tester
Share
Quick answer
Hanna salinity tester calibration takes a few steps: rinse and dry the tester, press CAL, cut open a fresh calibration sachet, put the electrodes fully in the solution, and tap out bubbles. The tester spots the solution on its own and shows "Stor" when it saves. This guide covers the Hanna HI98319 Marine Salinity Tester.
| Who | Reef keepers with a Hanna HI98319 Marine Salinity Tester. |
|---|---|
| What | A one-point or two-point calibration with Hanna's 35.00 ppt solution, and the 5.00 ppt solution if you use both. |
| When | Hanna says at least once a month with regular use, and any time you need high accuracy. |
| Where | At your bench, with the tester dry and the solution in a beaker or its sachet. |
| Why | Hanna says calibration is only as good as the solution used, so a fresh standard each time matters. |
Before you start
- The probe ships dry. Take off the protective cap before you use it.
- Rinse the tester with purified water and dry it completely.
- Use a fresh calibration standard each time. Hanna says the values of calibration solutions change over time.
- Hanna's product page lists two 35.00 ppt sachets and two 5.00 ppt sachets in the box. An older manual (10/18) only describes the 35.00 ppt point. This guide follows the newer manual (11/22).
Hanna salinity tester calibration, step by step
- Enter calibration mode. Press the CAL button. The tester shows "35.00 PPt USE" with the CAL tag blinking. If you set the range to Low in Setup, it shows "5.00 PPt USE" instead.
- Open the sachet. Cut the sachet open at the top. Push the edges to form a spout. Do not squeeze the sachet or warm the solution with your hands. Hanna suggests emptying it into a calibration beaker.
- Put the tester in. Place the tester in the solution so the electrodes are fully immersed.
- Tap out the bubbles. Tap the beaker to knock loose any trapped air.
- Wait for "Stor". The tester recognizes the solution by itself. "REC" shows until the reading is stable and accepted. Then the tester shows "Stor" while it saves the point and goes back to measuring mode.
To leave calibration without saving, press the CAL button. If the tester shows "Err" after three dashes, the reading was outside the accepted range for that solution.
One-point or two-point?
With the range set to High or Low in Setup, the tester supports one-point calibration. The default Auto range mode allows two-point calibration, and you follow the on-screen prompts. Bulk Reef Supply's walkthrough shows the single 35 ppt point only.
How to reset a calibration
Press CAL to enter calibration mode, then press ON/OFF. "CLR" shows, and the tester goes back to its default calibration.
Take a reading after calibrating
- Immerse. Put the tester 1.5 inches (38 mm) into the sample.
- Swirl. Swirl it to dislodge air bubbles and wait for the stability tag to go away. The tester adjusts for temperature on its own.
- Read. Salinity shows in the last unit you picked: ppt, PSU or S.G. The temperature shows on the second line.
- Rinse and cap. Rinse the probe with purified water, dry it, and put the protective cap back on.
Hanna's product page lists a range of 0.0 to 70.0 ppt and an accuracy of plus or minus 1 ppt up to 40.0 ppt. To switch units, open the battery cover and press the Setup button next to the battery.
How often to calibrate
Hanna's manual says to calibrate at least once a month with regular use. One reef retailer article says salinity meters need calibrating less often than pH meters. Follow Hanna's manual for this tester. Bulk Reef Supply also advises a plastic container, not glass or metal, and throwing the used solution away.
Care between calibrations
- Check the openings that hold the electrodes for foreign material. Hanna says you can clean them with a non-abrasive detergent and a soft material such as cardboard.
- After a deep clean, rinse in running tap water, rinse with purified water, dry the tester, and recalibrate before you use it.
- Store the tester with the protective cap on.
What salinity should you read?
Sources differ. Reef Builders (2020) gives 1.023 to 1.025 specific gravity. Bulk Reef Supply lists 35 ppt, or 1.026 specific gravity. Red Sea targets 33 to 35 PSU, depending on tank type.
Questions people ask
How often should I calibrate a Hanna salinity tester?
At least once a month with regular use, and when you need high accuracy, according to Hanna's manual.
Can I reuse calibration solution?
Hanna says to use a fresh standard for each calibration. Bulk Reef Supply says to discard used solution.
What does "Stor" mean?
The tester has accepted the calibration point and is saving it.
What does "CLR" mean?
The calibration is cleared and the tester is back on its default calibration.
Gear for this job
A home for your salinity tester
The Hanna Salinity Checker Stand is a 3D-printed holder made for Hanna Instruments' Salinity Tester, to store it when it is not in use. The 6-piece Hanna Checker Organization System covers more of your testing gear.
Shop Hanna checker organizersKeep reading: Reef Tank Parameters: The Numbers to Test and Target · Reef Tank Maintenance Schedule: Daily, Weekly, Monthly
Sources
- Hanna Instruments: HI98319 Marine Salinity Tester instruction manual (IST98319 11/22)
- Hanna Instruments: Marine Salinity Tester, HI98319
- Hanna Instruments: HI98319 instruction manual (older revision, IST98319 10/18)
- Bulk Reef Supply: How to Calibrate & Use the Hanna Instruments Marine Salinity Tester
- Bulk Reef Supply: Salinity Meter Calibration and How to Care for One
- Reef Builders: Important Water Parameters For Keeping A Saltwater Aquarium (2020)
- Bulk Reef Supply: A Weekly Saltwater Aquarium Maintenance Checklist
- Red Sea: Complete Reef Care Manual