How to Use a Hanna Alkalinity Checker
Share
Quick answer
To use a Hanna alkalinity checker, you zero it on 10 mL of plain tank water, add exactly 1 mL of reagent, invert the cuvette 5 times, and press the button again to read alkalinity in dKH. This guide covers the Hanna HI772 Marine Alkalinity Checker and follows Hanna's 02/22 manual.
| Who | Reef keepers with a Hanna HI772 Marine Alkalinity Checker HC. |
|---|---|
| What | A colorimetric test that reads alkalinity from 0.0 to 20.0 dKH in 0.1 dKH steps. |
| When | Reef Builders (2020) says weekly once a tank has cycled. Red Sea says every 2 to 3 weeks while you dose supplements. |
| Where | At your bench, with a clean, dry cuvette and a steady hand for the 1 mL of reagent. |
| Why | Alkalinity is one of the numbers that moves in a reef tank, and Red Sea sets a daily limit on how far to raise it. |
What you need
Hanna supplies these with the HI772: sample cuvette and cap (2), a marine alkalinity reagent starter kit for 25 tests, a 1 mL syringe with tip, a 1.5V AAA battery and the manual. You also need tank water. Hanna also sells an alkalinity checker that reads in ppm. This guide is for the HI772, which reads in dKH.
How to use a Hanna alkalinity checker, step by step
- Turn it on. Press the ON/OFF button. All segments show for a few seconds, then the display shows "Add", "C.1" with "Press" blinking.
- Fill the cuvette. Fill the cuvette with 10 mL of unreacted tank water and replace the cap. Wipe the outside dry, insert the cuvette into the checker, and close the cap.
- Zero the checker. Press ON/OFF. When the display shows "Add", "C.2" with "Press" blinking, the checker is zeroed.
- Add the reagent. Take the cuvette out and unscrew the cap. With the 1 mL syringe, carefully add exactly 1 mL of HI772S Marine Alkalinity reagent to the water. Replace the cap and gently invert 5 times.
- Read. Insert the cuvette, close the cap, and press ON/OFF. The display shows alkalinity in dKH. The checker turns off 10 minutes after the reading.
Do not spill reagent while you add it. Hanna warns that a spill can stop the full color from developing. The reaction runs through a range of colors from yellow to greenish blue.
Reading the result
The display is in dKH. Hanna gives the conversion: 1 dKH equals 17.86 ppm CaCO3, or 0.358 meq/L. If your supplement label uses ppm or meq/L, convert first. Accuracy is plus or minus 0.3 dKH plus or minus 5% of the reading at 25 C.
Tips for a better reading
- Keep debris out of the sample.
- The outside of the cuvette must be dry and free of fingerprints, oil and dirt. Wipe it with a microfiber cloth or a lint-free cloth before it goes in.
- Shaking makes bubbles, and bubbles raise the reading. Invert gently, and swirl or tap the cuvette to clear any bubbles.
- Use the same cuvette for the zero and the measurement.
- Do not let the reacted sample stand too long after you add reagent.
- Pour the sample out right after the reading. The glass can stain for good.
- Bulk Reef Supply suggests rinsing the vials with RO water before and after each use.
Once rinsed, cuvettes need somewhere to dry. The CuveDry Hanna cuvette drying rack holds each cuvette upright for air drying.
What the error messages mean
- Light High: too much ambient light reaches the detector. Check how you prepared the zero cuvette.
- Light Low: not enough light for a reading. Check the zero cuvette.
- Inverted Cuvettes: the sample and zero cuvettes are swapped. Swap them back and repeat.
- Under Range: a blinking "0.0" means the sample reads below the zero reference. Use the same cuvette for zero and measurement.
- Over Range: the top value blinks. Check for debris, dilute the sample, and repeat.
- Battery Low: replace the battery. A fresh battery lasts for at least 5000 measurements, Hanna says.
What alkalinity should you aim for?
Most sources we opened say 8 to 12 dKH (Reef Builders 2020, Bulk Reef Supply). Reef Builders (2008) says 8 to 11 dKH. Red Sea's table runs from 7 to 12 dKH, depending on tank type and goal. Red Sea also says to raise alkalinity by no more than 1.4 dKH in a day, to prevent stress to the corals.
Questions people ask
How much reagent does the HI772 use?
Exactly 1 mL of HI772S reagent, added to 10 mL of tank water with the 1 mL syringe.
What does "Add C.2" mean?
The checker has zeroed. Take the cuvette out and add the reagent.
How many times do I invert the cuvette?
Gently invert it 5 times after you add the reagent and replace the cap.
What is the range of the HI772?
0.0 to 20.0 dKH, with 0.1 dKH resolution.
Gear for this job
Keep the alkalinity kit in order
The Hanna Alkalinity Checker Kit is a 3D-printed organizer custom designed to fit the components of the Hanna Instruments Alkalinity Test kit. CuveDry holds Hanna Checker cuvettes upright to air dry.
Shop Hanna checker organizersKeep reading: Reef Tank Parameters: The Numbers to Test and Target · Hanna Calcium Checker Instructions, Step by Step
Sources
- Hanna Instruments: HI772 Marine Alkalinity instruction manual (IST772 02/22)
- Hanna Instruments: Saltwater Aquarium Calcium Colorimeter Checker HC, HI758 (lists the HI755 and HI772 alkalinity checkers)
- Bulk Reef Supply: Alkalinity Testing With Hanna's HI772 DKH Checker
- Reef Builders: Important Water Parameters For Keeping A Saltwater Aquarium (2020)
- Reef Builders: Stable Reef water parameters make a successful reef (2008)
- Bulk Reef Supply: A Weekly Saltwater Aquarium Maintenance Checklist
- Red Sea: Complete Reef Care Manual